PROGRAMMES:
Programmes | Temperature | Pressure | Sterilisation time | Time total | Type | Max load | Max load per tray |
Solid | 134 ºC | 210 kPa | 4 min | 20-35 min | Solid unwrapped material | 4.5 Kg | 1.2 Kg |
121 ºC | 110 kPa | 20 min | 30-35 min | ||||
Liquid | 134 ºC | 210 kPa | 4 min | 35-50 min | Liquid | 1 kg | 0.3 kg |
121 ºC | 110 kPa | 20 min | 35-55 min | ||||
Drying | - | - | - | 1-20 min | - | - | - |
- The minimum time required for the autoclave to be ready for use after connecting to power is 5 min.
- The maximum sterilisation cycle temperature of 134 ºC is 137 ºC
- The maximum sterilisation cycle temperature of 121 ºC is 124 ºC
FEATURES:
- Fully automatic pre-programmed sterilisation cycles.
- Chamber and door made of stainless steel with electro-polished finish.
- High precision control system for perfect sterilisation results.
- User friendly interface.
- LCD screen displays continously information about the cycle process.
- Open type water tank, which makes water filling and cleaning of the tank easier.
- Drying cycle.
- Optional external printer (not included) Ref: 922-PRINTER04.
- Languages: Spanish, English, German, French, Italian, Dutch, Polish, Romanian, Czech, Latvian, Lithuanian, Hungarian and Chinese.
SAFETY DEVICES:
Breaker.
Protect your computer from any possibility of failure in the power supply.
Action: Interrupt the power supply.
Thermal circuit breaker.
Protection against short circuits and possible overheating in the transformer primary network.
Action: Temporary interruption (until the temperature has dropped) in the winding.
Safety valve.
Protection against a possible increase in pressure within the chamber.
Action: Escape steam until the restoration of a normal pressure level.
Micro switch to control the state of the door.
Exercises control in adjusting the door to start a sterilization cycle.
Action: Transmits signal to alert the incorrect position of the door.
Thermostat for heating elements in the steam generator.
Protection against possible overheating in the heating elements in the steam generator.
Action: interrupts the power supply to the heating elements in the steam generator.
Thermostat for heating elements in the chamber.
Protection against possible overheating in the heating elements in the chamber.
Action: interrupts the power supply to the heating elements in the chamber.
Auto leveling hydraulic system.
The hydraulic system pressure returns to its natural if the sterilization cycle is interrupted manually.
Action: the internal pressure is restored automatically.
Door Lock
Effective to prevent accidental opening of the door protection.
Action: Prevents door opening by a hydraulic locking system.
Technical features:
- Nominal Voltage: 220-240V; 50/60 Hz; 10A
- Chamber dimensions: 247 x 350 mm
- Breaker: F16A / 400V
- Nominal power: 2200VA
- Sterilisation temperature: 121 ° C / 134 ° C
- Tank Capacity distilled water: 2.5 L approx.
- Operating temperature: 5 ° C / 40 ° C
- Operating Relative Humidity: Max. 80%, non-condensing
- External dimensions: 480 x 455 x 600 mm
- Net Weight: 45 Kg
- Level Max. Noise: <70 dB
- Operating atmospheric pressure: 76/106 kPa
COMPONENTS:
- Distilled water tank
- LCD display
- Control Panel
- Main Switch
- Key drain distilled and used water
- USB port
- External printer port
- Safety valve
- Ventilation grid
- Lock system
- Power Cable
- Label
ACCESSORIES INCLUDED:
- Steriliser
- Three instrument trays
- Base for trays
- Tray extraction handle
- Tool to adjust the door
- 2 drain hoses
- Door gasket
- Instruction manual
AUTOCLAVE PROCESS STEP BY STEP:
1- Open the door, take the tools and accessories that are within the sterilisation chamber, remove the wrapper and clean them.
2- Connect the power cord to a suitable power source.
3- Connect the printer.
4- Switch on: The main switch is located on the right bottom of the unit.
Once activated, the LCD screen lights, where basic operation icons of the steriliser are located, as the position of the door, the water level, the sterilisation programme, date, time, etc.
5- Fill tank with distilled water.
Remove the top lid and fill the tank, if you hear a signal, it means you exceeded the permitted level of water in the tank, please stop immediately and drain the amount of water necessary for the signal off.
6- Preparation of material to be sterilised:
Place instruments that are of different materials in different trays, however take care to separate them before sterilisation.
If the instruments are not stainless steel, put some sterilisation paper between the tray and the instruments.
Do not allow direct contact between instruments of different materials.
Verify that all instruments are in correct position.
Place on tray instruments such as lenses, containers, tubes, etc. inverted, so that water does not stagnate within them.
Do not overload the trays above their limit.
Do not overlap the trays on top of each other, or put them in direct contact with the wall of the sterilisation chamber.
Do not use metal clips or pins, this may impair the operation of the steriliser.
7- Select the sterilisation programme.
8- Start the sterilisation cycle.
After selecting the sterilisation programme, insert instruments into the chamber through the trays, using the supplied plastic handle.
Once the instruments are within the chamber, close the door.
Caution: Be sure to close the door, otherwise the steriliser displays an alarm and can not start or stop the programme during the cycle.
9. Start the sterilisation programme.
Press the start button, then the steriliser start operating during approximately between 30-75 minutes.
10-After the sterilisation cycle.
After completion of the sterilisation cycle, the printer will print the full report automatically.
After the pressure drops to zero, you can open the door and remove the instruments. The steriliser alers by a beep once the sterilisation cycle is completed.
Caution: Whenever you remove instruments, use the supplied plastic handle for supporting trays.